Each card represents a characteristic strain pattern for a specific disease. Clicking a card loads representative 17-segment values into the interactive bull's-eye. Use these as teaching references or to compare against your patient's actual data.
Normal ≤ −20%
Recommended ≥ 40 fps
Vendor-specific normal ranges (ASE/EACVI 2022)
| Vendor | Normal GLS (mean ± SD) | Lower Limit of Normal |
|---|---|---|
| GE HealthCare (EchoPAC) | −20.2 ± 1.9% | −16.4% |
| Philips (QLAB) | −20.4 ± 2.1% | −16.2% |
| Siemens (syngo.via) | −19.8 ± 2.0% | −15.8% |
| Canon Medical (Aplio) | −20.0 ± 2.0% | −16.0% |
| Fujifilm (Arietta) | −20.1 ± 2.0% | −16.1% |
| Vendor-neutral consensus | −20.0 ± 2.0% | −16.0% |
Normal mid-wall GLS is ≥ −17% (less negative than endocardial GLS). Mid-wall strain is more sensitive for subendocardial ischemia and early HCM. Most vendors report this automatically — check your software version for availability.
17-Segment Bull's-Eye
Click a segment to enter its strain value. Outer ring = basal, inner ring = apical.
17-Segment Strain Curves
Individual time-strain waveforms · Dashed = basal · Dotted = mid · Solid = apical/apex